Measurements on 3D models of human skulls derived from two different cone beam CT scanners

Measurements on 3D models of human skulls derived from two different cone beam CT scanners

  • نوع فایل : کتاب
  • زبان : انگلیسی
  • مؤلف : Olivier J. C. van Vlijmen & Frits A. Rangel & Stefaan J. Bergé & Ewald M. Bronkhorst & Alfred G. Becking & Anne Marie Kuijpers-Jagtman
  • چاپ و سال / کشور: 2010

Description

The aims of this study were to compare measurements on three-dimensional (3D) models of human skulls derived from two different cone beam CT scanners (CBCT) and to evaluate if the used hardware can influence the performed measurements. CBCT scans of 40 dry human skulls with both the i-CAT™ and the Iluma™ cone beam CT scanners were made. From the CBCT scans, 3D models were constructed. One operator identified 19 landmarks five times on both types of 3D models with a time interval of 1 week. Intra-observer reliability was high for most measurements. There was a statistically significant and clinically relevant difference for some measurements between constructed 3D models of the same skull from the two different CBCT devices. Used hardware for scanning might influence the measurements performed. This means that care should be taken when interpreting measurements made on CBCT 3D models derived from different CBCT device
Clin Oral Invest DOI 10.1007/s00784-010-0440-8 Received: 30 December 2009 / Accepted: 28 June 2010
اگر شما نسبت به این اثر یا عنوان محق هستید، لطفا از طریق "بخش تماس با ما" با ما تماس بگیرید و برای اطلاعات بیشتر، صفحه قوانین و مقررات را مطالعه نمایید.

دیدگاه کاربران


لطفا در این قسمت فقط نظر شخصی در مورد این عنوان را وارد نمایید و در صورتیکه مشکلی با دانلود یا استفاده از این فایل دارید در صفحه کاربری تیکت ثبت کنید.

بارگزاری