Multidetector CT imaging of mechanical prosthetic heart valves: quantification of artifacts with a pulsatile in-vitro model
- نوع فایل : کتاب
- زبان : انگلیسی
- مؤلف : Petr Symersky & Ricardo P. J. Budde & Paul Westers & Bas A. J. M. de Mol & Mathias Prokop
- چاپ و سال / کشور: 2011
Description
Objectives Multidetector computed tomography (MDCT) can detect the cause of prosthetic heart valve (PHV) dysfunction but is hampered by valve-induced artifacts. We quantified artifacts of four PHV using a pulsatile invitro model and assessed the relation to leaflet motion and valve design. Methods A Medtronic Hall tilting disc (MH), and Carbomedics (CM), St Jude (SJM), and ON-X bileaflet valves underwent CT in an in-vitro model using retrospective gating with a 64 detector CT system in stationary and pulsatile conditions. Artifacts and radiopaque component volumes were quantified with thresholds based on surrounding structures and valvular components. Results Hypodense artifacts volumes (mm3) were 1,029± 147, 535±53, 371±16, and 366±18 for the SJM, MH, CM and ON-X valves (p<0.001 except for the latter two valves p=0.43). Hyperdense artifact volumes were 3,546± 141, 2,387±103, 2,003±102, and 3,033±31 for the SJM, MH, CM and ON-X valve, respectively (all differences p< 0.001). Leaflet motion affected hypodense (F=41.5, p< 0.001) and hyperdense artifacts (F=53.7, p<0.001). Closed and moving leaflets were associated with the least and the most artifacts respectively (p<0.001, both artifact types). Conclusion Both valve design and leaflet motion affect PHV-induced artifacts. Best imaging results may be expected for the CM valve during phases in which the leaflets are closed.
Eur Radiol DOI 10.1007/s00330-011-2146-y Received: 2 March 2011 / Revised: 15 April 2011 / Accepted: 27 April 2011